Residual Chlorine Analyzer On-Line System
Negotiable Min Order Quantity Unit
- Required Quantity
-
- Place of Origin
- Payment Terms
- Negotiable
- Production method
- Negotiable
- Shipping / Lead Time
- Negotiable / Negotiable
- Keyword
- Category
- Water Treatment
Daeyoon Scale Industrial Co., Ltd.
- Country / Year Established
- South Korea /
- Business type
- Others
- Verified Certificate
-
16
Product name | Residual Chlorine Analyzer On-Line System | Certification | - |
---|---|---|---|
Category | Water Treatment | Ingredients | - |
Keyword | - | Unit Size | - |
Brand name | - | Unit Weigh | - |
origin | Stock | - | |
Supply type | - | HS code | - |
Product Information
= DWA-3000A RC Analyzer ( Real Time TREND ) =
Function / Feature :
â Real Time TREND Analyzer Function
â Simple Operation by 8 Keypad Button
â Maximum 4-Relay contact - HH, H, L, LL
â Large Size of Graphical LCD & backlight
â RS-232C/DC Power Modem Interface Communication
â Isolated Output 0~20, 0~24, 4~20mA Selection DC Output
â Isolated DC/AC Output Transmission
â Water-jet, Air-jet, Ultrasonic Wave, and Medicinal fluid Cleaning
of relay contact output
â Automatic/ Manual temperature compensation
â Automatic/ Manual calibration.
â Automatic cleaning control Function
â LOW-COST for user satisfaction
Application :
- Water Treatment Plant
- Pool,Beverage Process
- Industrial Process: Food/Chemical/Pharmaceutical etc.
- Disinfection Treatment
Specification :
Model Name ; DWA-3000A RC Analyzer
Measurement Range ; 0~2 ppm,(mg/â) or User Spec?
Power ; AC100V/AC220V 50/60Hz ±10% 3VA
Display ; Measurement Indication, Message Icon, Real time Trend Analysis Graph
Keypad ; Membrane Keypad 8 Keys
Temperature Range ; Temperature : -25°C ~ 70°C
Relative Humidity : Max. 90%(Non-condensing)
Accuracy ; ±1% of Full Scale
Repeatability Reproducibility ; ±1% of Full Scale
Response ; 120 sec (98% saturation)
Relay Output ; High-High, High, Low, Low-Low(4 Outputs)
Signal Output ; DC : Isolated 0~20, 0~24, 4~20mA_Option
Digital Interface : Isolated RS232C/DC POWER MODEM_Option
Condition ; Conductivity : Min. Over 100ã²/cm
pH : 6~8pH
Measurement Flow : Minimum 0~1 â/min.
Sample Water Temperature : 0~40°C
Sensor Calibration ; Auto/Manual Temperature Compensation
Calibration Method ; Colorimetry Method by Colorimeter
Dimensions ; 240(W) ?180(H) ?115(D)mm
Environmental Rating ; IP65, Waterproof, Dustproof, Outdoor installation structure
Installation ; 2″Pipe Mounting, Panel Wall Mounting
Weight ; About 1.5Kg
B2B Trade
Price (FOB) | Negotiable | transportation | - |
---|---|---|---|
MOQ | Negotiable | Leadtime | Negotiable |
Payment Options | Negotiable | Shipping time | Negotiable |
- President
- In-ho, Seo
- Address
- #207, Ace Techno Tower II, 197-7 Guro-3dong, Guro-gu, Seoul
- Product Category
- Home & Kitchen,Measuring & Gauging Tools,Other Cookware & Cooking Tools
- Company introduction
-
Our company was founded in 1982 and is specialized in electrochemical sensors and industrial measuring instruments. We aim to maximize our company value through wide experience and the management for customer satisfaction as a leader in the field.
Our company is producing high quality products through strict inspection.If you are interested in our products or have any questions, feel free to contact us.
- Main Markets
-
China
France
Germany
Singapore
Taiwan
- Main Product